| | | | | | | . | PHOPML: Basement Precision Measurement Lab Equipment Maintenance Downtime Log | | | | | | . | | | | | | | . | Color Code Definitions | DOWN - Equipment not usable | | | | | . | | UP - Equipment is operational | | | | | . | | Partially Operational - Equipment may be usable contact Lab Manager | | | | | . | | | | | | | . | Date | Equipment | Status | Follow on Notes | System Update | | . | 10/08/07 | Zeiss Supra 40VP | DOWN | No signals coming in when EHT turned on | | | . | 10/09/07 | Zeiss Supra 40VP | UP | Zeiss was brought to partially shut down and reboot | Service Complete | | . | 10/16/07 | Zeiss Supra 40VP | DOWN | stage not responding | | | . | 10/17/07 | Zeiss Supra 40VP | UP | service to clean column and restart computer and launch software | service complete | | . | 10/22/07 | Zeiss Supra 40VP | down | joystick pads doesn't respond, waiting for service | | | . | 10/23/07 | Zeiss Supra 40VP | UP | problem fixed, Chamber cleaned | Service Complete | | . | 10/23/2007 | Veeco Dimension 3000 | DOWN | computer reformatting to Windows XP need calibration data from vender. It should be back up early next week. | Software Upgrade | | . | 10/29/07 | Veeco Dimension 3000 | UP | Windows XP installed, USB works, afm caliberated | Maintenance Complete | | . | 11/01/07 | Zeiss Supra 40VP | DOWN | Trouble shooting for intermittent computer and instrument shut down. Column and chmaber were cleaned, and EHT was left on for 4 dyas. It should be back functioning after Monday. | Service in process | | . | 11/05/07 | Zeiss Supra 40VP | UP | Zeiss is functional (No problem occuured during the 4 days experiment)
| Service Complete | | . | 11/10/07 | Zeiss Supra 40VP | DOWN | alarm triggered on instrument, it's not functional to use | waiting for service | | . | 11/14/07 | Zeiss Supra 40VP | UP | carbon paint on bottom of sample holder resulted in shortage and stage lockup | Service Complete | | . | 11/20/07 | Zeiss Supra 40VP | Partially Operational | Knob sets on keyboard not responding, has to use mouse to do the work 11/27/07, either vacuum board or keyboard is malfunctioning, diagnose in progress, and watiting for parts | Service in process | |
|
| | | | | | | . | 12/03/07 | Zeiss Supra 40VP | UP | Hard panel keyboard problem fixed | Service Complete | | . | 12/21/07 | Zeiss Supra 40VP | DOWN | computer freezes, "initializing vacuum DLL" | waiting for repair | | . | 01/02/08 | Zeiss Supra 40VP | Partially Operational | PC replaced with new version of software, intermittent lock-up still present, waiting for replacement of final lens | Parts on Order | | . | 1/14/2008 | Zeiss Supra 40VP | DOWN | final lens is scheduled to be installed beginning Monday, January 14, at noon | Service in process | | . | 01/17/08 | Zeiss Supra 40VP | UP | final lens is replaced. | Service Complete | | . | 02/05/08 | Zeiss Supra 40VP e-beam writing | DOWN | e-beam writing controller needs to be replaced | Parts on Order | | . | 02/26/08 | Zeiss Supra 40VP e-beam writing | up | e-beam writing controllerwas replaced | Service Complete | | . | 06/13/08 | JEOL 6100 imaging | DOWN | not functioning for imaging, no focus control, etc. | Service Call Completed | | . | 06/19/08 | JEOL 6100 imaging | UP | | Service Call Completed | | . | 07/11/08 | JEOL 6100 imaging | DOWN | probe current knob doesn't function, monitor display doesn’t fuction | Service in process | | . | 07/18/08 | JEOL 6100 imaging | UP | | Service Call Completed | | . | 09/12/08 | Zygo New View 6300 basic | DOWN | computer is down | Service in process | | . | 09/18/08 | Zygo New View 6300 basic | UP | replaced power supply under Dell warrenty, also contacted Zygo tech support for software problem. | Service Call Completed | | . | 10/16/08 | JEOL 6100 imaging | DOWN | vacuum not efficient | Service in process | | . | 10/17/08 | JEOL 6100 imaging | UP | diffusion pump heater was replaced | Service Call Completed | | . | 11/15/08 | Zeiss Supra 40VP | DOWN | display doesn't work | Service in process | | . | 11/18/08 | Zeiss Supra 40VP | UP | Windows XP needed repair | Service Call Completed | | . | 11/17/08 | JEOL 6100 imaging | DOWN | fast scan doesn't work | Service in process | | . | 11/21/08 | JEOL 6100 imaging | UP | contacts are cleaned | Service Call Completed | | . | 01/30/09 | Zeiss Supra 40VP | DOWN | knob sets don't work, stage not function | Service in process | |
|
| | | | | | | . | 2/2/2009
| Zeiss Supra 40VP | Partially Operational | stage was reset and hard panel was replaced, but images won't get exported | Service in process | | . | 2/3/2009
| Zeiss Supra 40VP | UP | image problem was addressed | Service Call Completed | | . | 2/5/2009
| JEOL 6100 imaging | DOWN | TV mode doesn't work | waiting for service | | . | 2/11/09
| JEOL 6100 imaging | UP | Relay replaced, and it works | Service Call Completed | | . | 2/13/09
| JEOL 6100 imaging | Partially Operational | TV mode doesn't work | waiting for service | | . | 2/20/2009
2/24/09 | JEOL 6100 imaging | UP | replaced CRT and power supply, still waiting for parts to complete power supplly repair. Parts being replaced. Waiting for more | Parts on Order | | . | 03/09/09 | JEOL 6100 imaging | DOWN | Parts brought back, replaced, and needs more parts | Service in process | | . | | | | | | | . | 03/11/09 | JEOL 6100 imaging | up
| JEOL is functional , waiting for more parts | Parts on order | | . | 03/13/09 | JEOL 6100 | DOWN | "Board" was replaced to address overheating issue, but still not functioning | Service in progress | | . | 03/18/09 | JEOL 6100 | UP | | Service Call completed | | . | 05/04/09 | Zeiss Supra 40VP | Partially Operational | Rotation of the stage is not operational. Waiting for service and parts. | Service in progress | | . | 05/11/09 | JEOL 6100 | DOWN | Problems with imagining. Service has been called. | Service in progress | | . | 05/13/09 | JEOL 6100 | UP | problems was addressed | Service call completed | | . | 05/18/09 | JEOL 6100 | Down | TV mode blurry images, mag & scale bar not shown sometimes | Service call made
| | . | 05/04/09 | Zeiss Supra 40VP | UP | Rotation of the stage is working. Chiller rattle was fixed. | Service Completed | | . | 06/05/09 | Zeiss Supra 40VP | Partially operational | Hard Panel not functioning, no EBL communication Imaging working by using mouse to do all functions | Service in progress | | . | 06/12/09
| Zeiss Supra 40VP | Partially Operational | Hard Panel replaced, new version of SMT software installed | service in progress | | . | 6/23-24/2009 | Zeiss Supra 40VP | Down | repair of stage "lock-up", stage motor replaced, tension tape both directions replaced, stage board replaced, new hard drive installed | service in progress | | . | 6/25/2009
| Zeiss Supra 40VP | UP | | service completed | |
|
| | | | | | | . | 6/5/2009
| JEOL 6100 | down | TV mode not working | | | . | 6/8/2009
| JEOL 6100 | UP | Identified possible cause. Parts are being ordered. It's funtioning at the moment | parts on order | | . | 06/16/09 | JEOL 6100 | down | TV mode not working | service in progress
| | . | 7/19/2009
| Zeiss Supra 40VP | down | Stage doesn't move, and doesn' tinitialize | | | . | 7/20/2009
| Zeiss Supra 40VP | up | dirt and residues found on stage and moving device area | service completed | | . | 7/30/2009
| JEOL 6100 | up | Parts were replaced, and TV mode is working | service completed | | . | 8/28/2009
| Zeiss Supra 40VP | down | tip replace and vibration, air-leak issues afterwards | service in progress | | . | 9/2/2009
| Zeiss Supra 40VP | up | turbo pump replaced, air-leak noise fixed | service completed | | . | 10/23/09 | Veeco AFM Dimension | down | computer virus problem from memory stick | IT service in progress | | . | 10/29/09 | Veeco AFM Dimension | up | afm computer is working | service coompleted | | . | | | | | | | . | | | | | | | . | | | | | | | . | | | | | | | . | | | | | | | . | | | | | | | . | | | | | | | . | | | | | | | . | | | | | | | . | | | | | | |
|
|